Skip to main content
Log in

Laryngoscope burn risk in neonatal intubation

  • Short Communication
  • Published:
European Journal of Pediatrics Aims and scope Submit manuscript

Abstract

Laryngoscope burns in neonatal intubation

Following burns during neonatal intubation, we mounted an in vitro study of laryngoscopes to determine the temperatures reached during clinical use. The temperature of 10 different bulb laryngoscopes heads and two fibre optic heads were measured with a thermocouple, once opened, and upon closing. Within 60 s, all ten laryngoscopes, with light-bulb sources, had gained significant heat to cause thermal injury to neonatal skin. Laryngoscopes with LED light source and fibre optic heads did not.

Conclusion: We recommend that the bulb laryngoscope blade, if used, is not left open prior to intubation and that it is closed between intubation attempts.

What is Known:

The preterm epidermis is particularly vulnerable to injury.

What is New:

Bulb laryngoscope light bulbs consistently reach temperatures sufficient to burn neonatal skin in less than 100 s in an in vitro study.

Bulb light safety advice should be incorporated into intubation guidelines.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

References

  1. Grunau RE, Holsti L, Haley DW, Oberlander T, Weinberg J, Solimano A, Whitfield MF, Fitzgerald C, Yu W (2005) Neonatal procedural pain exposure predicts lower cortisol and behavioral reactivity in preterm infants in the NICU. Pain 113(3):293–300. https://doi.org/10.1016/j.pain.2004.10.020

    Article  PubMed  PubMed Central  Google Scholar 

  2. Koh TH, Coleman R (2000) Oropharyngeal burn in a newborn baby: new complication of light-bulb laryngoscopes. Anesthesiology 92(1):277–279

    Article  CAS  PubMed  Google Scholar 

  3. Lewis E, Zatman ST, Wilkes AR, Hall JE (2009) Laryngoscope light output. Anaesthesia 64(6):688–689. https://doi.org/10.1111/j.1365-2044.2009.05948.x

    Article  CAS  PubMed  Google Scholar 

  4. Mainous RO, Looney S (2007) A pilot study of changes in cerebral blood flow velocity, resistance, and vital signs following a painful stimulus in the premature infant. Adv Neonatal Care 7(2):88–104

    Article  PubMed  Google Scholar 

  5. Moritz AR, Henriques FC (1947) Studies of thermal injury: II. The relative importance of time and surface temperature in the causation of cutaneous burns. Am J Pathol 23(5):695–720

    CAS  PubMed  PubMed Central  Google Scholar 

  6. Toung TJ, Donham RT, Shipley R (1981) Thermal burn caused by a laryngoscope. Anesthesiology 55(2):184–185

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Contributions

ER drafted the study. All authors were involved in the design of the experiment. PON performed the experiment and recorded the results under DB’s supervision. ER wrote the final manuscript. JFA critically revised and approved the manuscript.

Corresponding author

Correspondence to Emer Ryan.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Research involving human participants/and or animals

“This article does not contain any studies with human participants or animals performed by any of the authors.”

Consent

No individual participants included in the study.”

Additional information

Communicated by Patrick Van Reempts

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ryan, E., O Neill, P., Brabazon, D. et al. Laryngoscope burn risk in neonatal intubation. Eur J Pediatr 178, 1125–1127 (2019). https://doi.org/10.1007/s00431-019-03381-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00431-019-03381-0